Changchun Institute of Optics,Fine Mechanics and Physics,CAS
Improved measurement accuracy of spot position on an InGaAs quadrant detector | |
Wu, J. B.; Y. S. Chen; S. J. Gao; Y. M. Li and Z. Y. Wu | |
2015 | |
发表期刊 | Applied Optics |
卷号 | 54期号:27页码:8049-8054 |
摘要 | In this paper, a new formula is proposed to improve the accuracy of spot position measurements on an InGaAs quadrant detector (QD). It is obtained by analyzing the relationship between the light spot position and the output signals of the QD and combining the infinite integral method with the Boltzmann method due to their opposite error characteristics. Based on the proposed formula, the measurement accuracy can be improved greatly, which is confirmed by the simulation and experimental results. In addition, it requires fewer parameters compared with the polynomial method when reaching the same accuracy. Thus, the new formula can be practical in applications of spot position measurements. (C) 2015 Optical Society of America |
收录类别 | SCI |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/55285 |
专题 | 中科院长春光机所知识产出 |
推荐引用方式 GB/T 7714 | Wu, J. B.,Y. S. Chen,S. J. Gao,et al. Improved measurement accuracy of spot position on an InGaAs quadrant detector[J]. Applied Optics,2015,54(27):8049-8054. |
APA | Wu, J. B.,Y. S. Chen,S. J. Gao,&Y. M. Li and Z. Y. Wu.(2015).Improved measurement accuracy of spot position on an InGaAs quadrant detector.Applied Optics,54(27),8049-8054. |
MLA | Wu, J. B.,et al."Improved measurement accuracy of spot position on an InGaAs quadrant detector".Applied Optics 54.27(2015):8049-8054. |