Changchun Institute of Optics,Fine Mechanics and Physics,CAS
Absorptance measurement for sloping bottom cavity of cryogenic radiometer | |
Yi, X.-L.; Z.-L. Yang; X. Ye; K. Wang; W. Fang and Y.-P. Wang | |
2015 | |
发表期刊 | Guangxue Jingmi Gongcheng/Optics and Precision Engineering |
卷号 | 23期号:10页码:2733-2739 |
摘要 | The Total Solar Irradiance (TSI) cavity in an Absolute Radiance Calibration Primary Radiometer(ARCPR) for space remote sensing should have an absorptance more than 0.9999 and the measurement uncertainty superior to 0.001%. Therefore, a cylindrical sloping bottom cavity was used as a receiver cavity for calibration of the TSI because of its superior performance in the space and cryogenic environments. Then, the absorptance of the cylindrical sloping bottom cavity was measured. The characteristics of the cylindrical sloping bottom cavity were introduced, and why the cavity was used as the receiver cavity in cryogenic radiometer was analyzed. Furthermore, the principle of the absorptance measured by a substitution method was expounded. A reference light path was added to monitor the stability of a laser to improve the measurement repeatability and accuracy. Finally, the absorptance of sloping bottom cavity was measured using this method, and the uncertainty of measurement results was analyzed. Experimental results indicate that the absorptance of sloping bottom cavity is 0.999 9280.000 005, superior to that of the calibration standard of the blackbody cavity, which verifies it is feasibility to use the sloping bottom cavity as the TSI cavity for measuring the TSI. The experiments also verify that the ratio of single voltage and reference voltage could be used to calculate the cavity absorptance, and could improve the uncertainty of measurement result. This method is fit for measuring the absorptance of an ultra high absorptance cavity. 2015, SCIENCE PRESS. All right reserved. |
文章类型 | 期刊论文 |
收录类别 | EI |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/56127 |
专题 | 中科院长春光机所知识产出 |
推荐引用方式 GB/T 7714 | Yi, X.-L.,Z.-L. Yang,X. Ye,et al. Absorptance measurement for sloping bottom cavity of cryogenic radiometer[J]. Guangxue Jingmi Gongcheng/Optics and Precision Engineering,2015,23(10):2733-2739. |
APA | Yi, X.-L.,Z.-L. Yang,X. Ye,K. Wang,&W. Fang and Y.-P. Wang.(2015).Absorptance measurement for sloping bottom cavity of cryogenic radiometer.Guangxue Jingmi Gongcheng/Optics and Precision Engineering,23(10),2733-2739. |
MLA | Yi, X.-L.,et al."Absorptance measurement for sloping bottom cavity of cryogenic radiometer".Guangxue Jingmi Gongcheng/Optics and Precision Engineering 23.10(2015):2733-2739. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
低温辐射计斜底腔吸收比测量.caj(251KB) | 期刊论文 | 作者接受稿 | 开放获取 | CC BY-NC-SA | 浏览 下载 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论