摘要 | Both thermal sensitivity and mass sensitivity in liquid of the first antisymmetric (A(0)) mode and the first symmetric (S(0)) mode of Lamb wave biosensor were investigated. A(0) and S(0) modes are sensitive to the mass change on the surface of the sensor but A(0) mode is also sensitive to the liquid in the region of evanescent wave associated with Lamb wave. By combining A(0) mode and S(0) mode, the measurement error due to the environmental temperature drift decreased by a large factor, therefore, the environmental temperature was efficiently compensated without changing the structure of Lamb wave sensor. (c) 2008 American Institute of Physics. |
收录类别 | SCI ; EI |
语种 | 英语 |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/26485 |
专题 | 中科院长春光机所知识产出 |
推荐引用方式 GB/T 7714 | Li F.,Wu Y. H.,Manceau J. F.,et al. Temperature compensation of lamb wave sensor by combined antisymmetric mode and symmetric mode[J]. Applied Physics Letters,2008,92(7). |
APA | Li F.,Wu Y. H.,Manceau J. F.,&Bastien F..(2008).Temperature compensation of lamb wave sensor by combined antisymmetric mode and symmetric mode.Applied Physics Letters,92(7). |
MLA | Li F.,et al."Temperature compensation of lamb wave sensor by combined antisymmetric mode and symmetric mode".Applied Physics Letters 92.7(2008). |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Li-2008-Temperature (585KB) | 开放获取 | -- | 浏览 下载 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论