Changchun Institute of Optics,Fine Mechanics and Physics,CAS
Analysis and Correction of Polarization Response Calibration Error of Limb Atmosphere Ultraviolet Hyperspectral Detector | |
H. C. Li; Z. F. Li; Y. Huang; G. Y. Lin; J. X. Zeng; H. S. Li; S. R. Wang and W. Y. Han | |
2022 | |
发表期刊 | Sensors |
卷号 | 22期号:21页码:20 |
摘要 | A UV hyperspectral instrument was designed with a polarization measurement channel for real-time in-orbit polarization correction to reduce the influence of polarization on the detection accuracy of atmospheric radiation. One of the prerequisites for in-orbit polarization calibration is accurately calibrating the instrument's polarization properties in the laboratory. This study first introduces the calibration method and measuring device of the polarization characteristics of the ultraviolet (UV) hyperspectral detector and conducts a polarization calibration test of the instrument. The two main error sources introduced by the calibration device were emphatically analyzed, and the correction method of the error sources was deduced theoretically. Finally, the polarization calibration results of the UV hyperspectral detector were corrected, and the uncertainty analysis of the corrected calibration results was about 1.4%, which provides effective ground polarization calibration data for the on-orbit polarization correction of the instrument. |
DOI | 10.3390/s22218542 |
URL | 查看原文 |
收录类别 | sci ; ei |
语种 | 英语 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/66409 |
专题 | 中国科学院长春光学精密机械与物理研究所 |
推荐引用方式 GB/T 7714 | H. C. Li,Z. F. Li,Y. Huang,et al. Analysis and Correction of Polarization Response Calibration Error of Limb Atmosphere Ultraviolet Hyperspectral Detector[J]. Sensors,2022,22(21):20. |
APA | H. C. Li.,Z. F. Li.,Y. Huang.,G. Y. Lin.,J. X. Zeng.,...&S. R. Wang and W. Y. Han.(2022).Analysis and Correction of Polarization Response Calibration Error of Limb Atmosphere Ultraviolet Hyperspectral Detector.Sensors,22(21),20. |
MLA | H. C. Li,et al."Analysis and Correction of Polarization Response Calibration Error of Limb Atmosphere Ultraviolet Hyperspectral Detector".Sensors 22.21(2022):20. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Analysis and Correct(6360KB) | 期刊论文 | 出版稿 | 开放获取 | CC BY-NC-SA | 浏览 下载 |