CIOMP OpenIR
Effects of a misaligned photodetector in autocollimators on angle measurements
J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang
2020
发表期刊Chinese Optics
ISSN20951531
卷号13期号:3页码:558-567
摘要As one of the key errors in autocollimators, misalignment of the photodetector is analyzed and modeled carefully in this paper. Effect of misalignment of the photodetector on angle measurements is characterized, when the photodetector in any position and orientation in space with respect to the theoretical image plane of autocollimators. It is shown that the angular measurement errors of autocollimators induced by a misaligned photodetector increase with greater measuring range L, larger angle and smaller focal length f of the collimating object lens. When f=300 mm, L=100 mm, =20, the angular measurement error caused by a misaligned photodetector is 0.004 5. The effects of each photodetector misalignment error on angle measurements in autocollimators are characterized. The model proposed in this paper is validated. Among all kinds of photodetector misalignment errors, the defocusing error has the greatest influences on autocollimators. Hence, it is critical to choose an imaging objective with longer focal length, reduce the measurement distance, and improve the installation accuracy of the photodetector along the axis. The model proposed in this paper helps to systematically obtain the angular measurement errors caused by a misaligned photodetector, which will play a key role in building a better error analysis model for autocollimators. 2020, China Science Publishing & Media LTD. All right reserved.
DOI10.3788/CO.2019-0207
URL查看原文
收录类别EI
引用统计
正在获取...
文献类型期刊论文
条目标识符http://ir.ciomp.ac.cn/handle/181722/64477
专题中国科学院长春光学精密机械与物理研究所
推荐引用方式
GB/T 7714
J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang. Effects of a misaligned photodetector in autocollimators on angle measurements[J]. Chinese Optics,2020,13(3):558-567.
APA J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang.(2020).Effects of a misaligned photodetector in autocollimators on angle measurements.Chinese Optics,13(3),558-567.
MLA J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang."Effects of a misaligned photodetector in autocollimators on angle measurements".Chinese Optics 13.3(2020):558-567.
条目包含的文件 下载所有文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Effects of a misalig(3472KB)期刊论文出版稿开放获取CC BY-NC-SA浏览 下载
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang]的文章
百度学术
百度学术中相似的文章
[J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang]的文章
必应学术
必应学术中相似的文章
[J. Luo,X.-H. Zhang,X. He,L. Ye and T.-Y. Zhang]的文章
相关权益政策
暂无数据
收藏/分享
文件名: Effects of a misaligned photodetector in autoc.pdf
格式: Adobe PDF
Get Adobe Flash player
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。