Body size and tube voltage dependent corrections for Hounsfield Unit in medical X-ray computed tomography: theory and experiments | |
X. M. Zheng, Y. Al-Hayek, C. Cummins, X. T. Li, L. Nardi, K. Albari, J. Evans, E. Roworth and T. Seaton | |
2020 | |
发表期刊 | Scientific Reports
![]() |
ISSN | 2045-2322 |
卷号 | 10期号:1页码:10 |
摘要 | The purpose of this work is to present a body size and tube voltage dependent correction scheme for the Hounsfield Unit, HU, in medical X-ray Computed Tomography imaging. Boltzmann photon transport equation was employed to study X-ray interaction with bulk water in CT imaging. Experimentally measured X-ray output in body of phantoms and attenuation cross sections of water were employed in the derivation of beam intensity in X-ray imaging. A Somatom Emotion CT scanner from Siemens and electron density phantoms from CIRS were employed to acquire CT images of different body sizes and different tissue materials located at different depths from body's surface. Tube voltage and depth dependent effective attenuation of bulk water was found from theoretical analysis in agreement with measured size-specific correction factors for-CTDIvol under different tube voltages. A size and tube voltage dependent correction scheme for the Hounsfield Unit is established. For the same tissue material, body size has much larger impact on the CT number variations than that of depth from the body surface in phantom measurements. Good results were achieved by applying the established correction scheme on the experimentally measured CT number variations under different tube voltages and body sizes. |
DOI | 10.1038/s41598-020-72707-y |
URL | 查看原文 |
收录类别 | SCI |
语种 | 英语 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/64336 |
专题 | 中国科学院长春光学精密机械与物理研究所 |
推荐引用方式 GB/T 7714 | X. M. Zheng, Y. Al-Hayek, C. Cummins, X. T. Li, L. Nardi, K. Albari, J. Evans, E. Roworth and T. Seaton. Body size and tube voltage dependent corrections for Hounsfield Unit in medical X-ray computed tomography: theory and experiments[J]. Scientific Reports,2020,10(1):10. |
APA | X. M. Zheng, Y. Al-Hayek, C. Cummins, X. T. Li, L. Nardi, K. Albari, J. Evans, E. Roworth and T. Seaton.(2020).Body size and tube voltage dependent corrections for Hounsfield Unit in medical X-ray computed tomography: theory and experiments.Scientific Reports,10(1),10. |
MLA | X. M. Zheng, Y. Al-Hayek, C. Cummins, X. T. Li, L. Nardi, K. Albari, J. Evans, E. Roworth and T. Seaton."Body size and tube voltage dependent corrections for Hounsfield Unit in medical X-ray computed tomography: theory and experiments".Scientific Reports 10.1(2020):10. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Zheng-2020-Body size(3860KB) | 期刊论文 | 出版稿 | 开放获取 | CC BY-NC-SA | 浏览 下载 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论