CIOMP OpenIR  > 中科院长春光机所知识产出
Calibration algorithm for cooled mid-infrared systems considering the influences of ambient temperature and integration time
S.T.Chang; Z.Li
2019
发表期刊Applied Optics
ISSN1559-128X
卷号58期号:29页码:8118-8125
摘要Midwave infrared systems with cooled detectors are generally used for high-precision or quantitative measurement, such as radiometry and thermometry. As a basis of these applications, radiometric calibration aims to obtain the relationship between the infrared images and the incident radiant flux generated by the scene or targets. Conventional radiometric calibration algorithms do not take the influences of integration and ambient temperature into consideration. As a consequence, the accuracy of calibration deteriorates whenever the temperature or the integration time varies. To solve this problem, we analyzed the effects of integration time and ambient temperature on coefficients of the radiometric calibration formula by theoretical and experimental analysis. Then, a radiometric calibration method is deduced to remove the variation of integration time and ambient temperature on the accuracy of calibration and radiometry. Several radiometric calibration experiments were conducted using a midwave infrared camera inside a chamber with controllable temperature. The results indicate that the proposed calibration algorithm is more effective and accurate, compared with conventional calibration methods, in complicated working conditions with variable integration times and ambient temperatures. (C) 2019 Optical Society of America
关键词nonuniformity correction,radiation,radiance,camera,gain,Optics
DOI10.1364/ao.58.008118
URL查看原文
收录类别SCI ; EI
语种英语
引用统计
文献类型期刊论文
条目标识符http://ir.ciomp.ac.cn/handle/181722/63467
专题中科院长春光机所知识产出
推荐引用方式
GB/T 7714
S.T.Chang,Z.Li. Calibration algorithm for cooled mid-infrared systems considering the influences of ambient temperature and integration time[J]. Applied Optics,2019,58(29):8118-8125.
APA S.T.Chang,&Z.Li.(2019).Calibration algorithm for cooled mid-infrared systems considering the influences of ambient temperature and integration time.Applied Optics,58(29),8118-8125.
MLA S.T.Chang,et al."Calibration algorithm for cooled mid-infrared systems considering the influences of ambient temperature and integration time".Applied Optics 58.29(2019):8118-8125.
条目包含的文件 下载所有文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Calibration algorith(924KB)期刊论文出版稿开放获取CC BY-NC-SA浏览 下载
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[S.T.Chang]的文章
[Z.Li]的文章
百度学术
百度学术中相似的文章
[S.T.Chang]的文章
[Z.Li]的文章
必应学术
必应学术中相似的文章
[S.T.Chang]的文章
[Z.Li]的文章
相关权益政策
暂无数据
收藏/分享
文件名: Calibration algorithm for cooled mid infrared.pdf
格式: Adobe PDF
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。