CIOMP OpenIR
Evaluation of different calibration equations for NTC thermistor applied to high-precision temperature measurement
Liu, G.; Guo, L.; Liu, C. L.; Wu, Q. W.
2018
发表期刊Measurement
ISSN0263-2241
卷号120页码:21-27
摘要Using a negative temperature coefficient (NTC) thermistor for high-precision temperature measurement can give a resolution and accuracy as low as 5 mK. The performance of an NTC thermistor is affected markedly by its calibration equation. Two series of high precision calibration for NTC thermistors in a precision water bath by means of comparison method were presented. Nine approximate calibration equations for the resistance-temperature characteristics of the MF501 NTC thermistor are evaluated within a temperature range of 278.15-328.15 K. It is confirmed that the fitting quality is influenced greatly by the number of coefficients used in the calibration equation, and that the Hoge-2 equation is the best calibration equation for the MF501 NTC thermistor for high-precision temperature measurements. The combined standard uncertainty of the thermistor calibration system is estimated as 4.31 mK. The calibration procedure and evaluation method proposed can be used for calibration of any types of NTC thermistors.
关键词Calibration equation NTC thermistor High precision Uncertainty Engineering Instruments & Instrumentation
DOI10.1016/j.measurement.2018.02.007
收录类别SCI ; EI
引用统计
文献类型期刊论文
条目标识符http://ir.ciomp.ac.cn/handle/181722/60982
专题中国科学院长春光学精密机械与物理研究所
推荐引用方式
GB/T 7714
Liu, G.,Guo, L.,Liu, C. L.,et al. Evaluation of different calibration equations for NTC thermistor applied to high-precision temperature measurement[J]. Measurement,2018,120:21-27.
APA Liu, G.,Guo, L.,Liu, C. L.,&Wu, Q. W..(2018).Evaluation of different calibration equations for NTC thermistor applied to high-precision temperature measurement.Measurement,120,21-27.
MLA Liu, G.,et al."Evaluation of different calibration equations for NTC thermistor applied to high-precision temperature measurement".Measurement 120(2018):21-27.
条目包含的文件 下载所有文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Evaluation of differ(495KB)期刊论文出版稿开放获取CC BY-NC-SA浏览 下载
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Liu, G.]的文章
[Guo, L.]的文章
[Liu, C. L.]的文章
百度学术
百度学术中相似的文章
[Liu, G.]的文章
[Guo, L.]的文章
[Liu, C. L.]的文章
必应学术
必应学术中相似的文章
[Liu, G.]的文章
[Guo, L.]的文章
[Liu, C. L.]的文章
相关权益政策
暂无数据
收藏/分享
文件名: Evaluation of different calibration equations.pdf
格式: Adobe PDF
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。