CIOMP OpenIR  > 中科院长春光机所知识产出
Time constant optimization of solar irradiance absolute radiometer
Tang, X.; W. Fang; Y.-p. Wang; D.-j. Yang and X.-l. Yi
2017
发表期刊Optoelectronics Letters
卷号13期号:3
摘要We experimentally evaluate and optimize the time constant of solar irradiance absolute radiometer (SIAR). The systemic error introduced by variable time constant is studied by a finite element method. The results shown that, with a classic time constant of 30 s for SIAR, the systemic errors are 0.06% in the midday and 0.275% in the morning and afternoon. The uncertainty level which can be considered negligible for SIAR is also investigated, and it is suggested that the uncertainty level has to be less than 0.02%. Then, combining the requirement of international comparison with these two conclusions, we conclude that the suitable time constant for SIAR is 20 s.1. 2017, Tianjin University of Technology and Springer-Verlag Berlin Heidelberg.
收录类别sci ; ei
语种英语
文献类型期刊论文
条目标识符http://ir.ciomp.ac.cn/handle/181722/59212
专题中科院长春光机所知识产出
推荐引用方式
GB/T 7714
Tang, X.,W. Fang,Y.-p. Wang,et al. Time constant optimization of solar irradiance absolute radiometer[J]. Optoelectronics Letters,2017,13(3).
APA Tang, X.,W. Fang,Y.-p. Wang,&D.-j. Yang and X.-l. Yi.(2017).Time constant optimization of solar irradiance absolute radiometer.Optoelectronics Letters,13(3).
MLA Tang, X.,et al."Time constant optimization of solar irradiance absolute radiometer".Optoelectronics Letters 13.3(2017).
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Time constant optimi(846KB)期刊论文作者接受稿开放获取CC BY-NC-SA浏览 请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Tang, X.]的文章
[W. Fang]的文章
[Y.-p. Wang]的文章
百度学术
百度学术中相似的文章
[Tang, X.]的文章
[W. Fang]的文章
[Y.-p. Wang]的文章
必应学术
必应学术中相似的文章
[Tang, X.]的文章
[W. Fang]的文章
[Y.-p. Wang]的文章
相关权益政策
暂无数据
收藏/分享
文件名: Time constant optimization.pdf
格式: Adobe PDF
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。