Changchun Institute of Optics,Fine Mechanics and Physics,CAS
A prior-based metal artifact reduction algorithm for x-ray CT | |
Li, M.![]() | |
2015 | |
发表期刊 | Journal of X-Ray Science and Technology
![]() |
卷号 | 23期号:2页码:229-241 |
摘要 | In computed tomography (CT), metal objects in the scanning filed are accompanied by physical phenomenon that causes projections to be inconsistent. These inconsistencies produce bright and dark shadows or streaks in analytically reconstructed images. Interpolation-based metal artifact reduction (MAR) algorithms usually replace the inconsistent projection data by estimating surrogate data based on the surrounding uncorrupted projections. In such cases, secondary artifacts will be generated when the data estimates are inaccurate. Therefore, better projection estimation is critical. This paper proposes an image post-processing strategy to create an intermediate image, named the prior image and better estimates of the surrogate data by forward projecting this prior image. The proposed method consists of three steps based on the forward projection MAR framework. First, metallic implants in the uncorrected images are segmented using a Markov random field model (MRF). Then a prior image is generated via an edge-preserving filter and a recovery procedure of the adjacent anatomical structures. Finally, the projection is completed via forward projecting this prior image and the corrected image is reconstructed by the filtered back-projection (FBP) method. Studies on both phantom and clinical data are carried out to verify the performance of the proposed method. The comparisons with other previous MAR algorithms demonstrate that the proposed MAR method performs better in metal artifact suppression and anatomical structure preservation. |
收录类别 | SCI ; EI |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/55302 |
专题 | 中科院长春光机所知识产出 |
推荐引用方式 GB/T 7714 | Li, M.,J. Zheng,T. Zhang,et al. A prior-based metal artifact reduction algorithm for x-ray CT[J]. Journal of X-Ray Science and Technology,2015,23(2):229-241. |
APA | Li, M.,J. Zheng,T. Zhang,Y. H. Guan,&P. Xu and M. S. Sun.(2015).A prior-based metal artifact reduction algorithm for x-ray CT.Journal of X-Ray Science and Technology,23(2),229-241. |
MLA | Li, M.,et al."A prior-based metal artifact reduction algorithm for x-ray CT".Journal of X-Ray Science and Technology 23.2(2015):229-241. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Li-2015-A prior-base(3420KB) | 期刊论文 | 作者接受稿 | 开放获取 | CC BY-NC-SA | 浏览 下载 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论