Changchun Institute of Optics,Fine Mechanics and Physics,CAS
Calibration of 1 m aperture infrared theodolite | |
Li N.; Zhang Y.-F.; Liu C.-X.; Cao L.-H.; Guo L.-H. | |
2014 | |
发表期刊 | Guangxue Jingmi Gongcheng/Optics and Precision Engineering |
ISSN | ISBN/1004924X |
卷号 | 22期号:8页码:2054-2060 |
摘要 | To measure infrared radiation characteristics of air targets, the calibration system for a 1 m aperture infrared measurement system was designed. A mathematic model of calibration for the infrared measurement system was given and a radiometric calibration system was established by combinations of a cavity blackbody and a collimator as the standard radiation source. Because the spectral response of the infrared measuring system is selective, the calibration principle based on conventional radiometric calibration method has theoretical errors and reduces the precision of radiometric calibration. To solve the problem, a relative spectral calibration method based on a monochromator and a standard radiometer was proposed. The mathematical model was given and a relative spectral calibration was performed. Finally, the experiments on radiation characteristic measurement were performed in the field by using the blackbody as a simulated target. The results show that the measurement max error from the 1 m aperture infrared radiation characteristic measurement system is 9.5% and improves by an average of 8.7% as compared with that of the conventional method. These show that the proposed method meets the requirements of practical applications and is suitable for the radiometric calibration in the out-fields. |
收录类别 | EI |
语种 | 中文 |
文献类型 | 期刊论文 |
条目标识符 | http://ir.ciomp.ac.cn/handle/181722/43777 |
专题 | 中科院长春光机所知识产出 |
推荐引用方式 GB/T 7714 | Li N.,Zhang Y.-F.,Liu C.-X.,et al. Calibration of 1 m aperture infrared theodolite[J]. Guangxue Jingmi Gongcheng/Optics and Precision Engineering,2014,22(8):2054-2060. |
APA | Li N.,Zhang Y.-F.,Liu C.-X.,Cao L.-H.,&Guo L.-H..(2014).Calibration of 1 m aperture infrared theodolite.Guangxue Jingmi Gongcheng/Optics and Precision Engineering,22(8),2054-2060. |
MLA | Li N.,et al."Calibration of 1 m aperture infrared theodolite".Guangxue Jingmi Gongcheng/Optics and Precision Engineering 22.8(2014):2054-2060. |